Jump to content

Gas chromatography

From Wikipedia, the free encyclopedia
Gas chromatography
A gas chromatograph with a headspace sampler
AcronymGC
ClassificationChromatography
Analytes
Other techniques
Related
HyphenatedGas chromatography-mass spectrometry

Gas chromatography (GC) is a common type of chromatography used in analytical chemistry for separating and analyzing compounds that can be vaporized without decomposition. Typical uses of GC include testing the purity of a particular substance, or separating the different components of a mixture.[1] In preparative chromatography, GC can be used to prepare pure compounds from a mixture.[2][3]

Gas chromatography is also sometimes known as vapor-phase chromatography (VPC), or gas–liquid partition chromatography (GLPC). These alternative names, as well as their respective abbreviations, are frequently used in scientific literature.[2]

Gas chromatography is the process of separating compounds in a mixture by injecting a gaseous or liquid sample into a mobile phase, typically called the carrier gas, and passing the gas through a stationary phase. The mobile phase is usually an inert gas or an unreactive gas such as helium, argon, nitrogen or hydrogen.[1] The stationary phase can be solid or liquid, although most GC systems today use a polymeric liquid stationary phase.[4] The stationary phase is contained inside of a separation column. Today, most GC columns are fused silica capillaries with an inner diameter of 100–320 micrometres (0.0039–0.0126 in) and a length of 5–60 metres (16–197 ft). The GC column is located inside an oven where the temperature of the gas can be controlled and the effluent coming off the column is monitored by a suitable detector.[1]

Operating principle

[edit]
Diagram of a gas chromatograph

A gas chromatograph is made of a narrow tube, known as the column, through which the vaporized sample passes, carried along by a continuous flow of inert or nonreactive gas. Components of the sample pass through the column at different rates, depending on their chemical and physical properties and the resulting interactions with the column lining or filling, called the stationary phase. The column is typically enclosed within a temperature controlled oven. As the chemicals exit the end of the column, they are detected and identified electronically.[1]

History

[edit]
Gas chromatograph

Background

[edit]

Chromatography dates to 1903 in the work of the Russian scientist, Mikhail Semenovich Tswett,[5] who separated plant pigments via liquid column chromatography.

Invention

[edit]

The invention of gas chromatography is generally attributed to Anthony T. James and Archer J.P. Martin.[6][7] Their gas chromatograph used partition chromatography as the separating principle, rather than adsorption chromatography. The popularity of gas chromatography quickly rose after the development of the flame ionization detector.[8] Martin and another one of their colleagues, Richard Synge, with whom he shared the 1952 Nobel Prize in Chemistry, had noted in an earlier paper[9] that chromatography might also be used to separate gases. Synge pursued other work while Martin continued his work with James.

Gas adsorption chromatography precursors

[edit]

German physical chemist Erika Cremer in 1947 together with Austrian graduate student Fritz Prior developed what could be considered the first gas chromatograph that consisted of a carrier gas, a column packed with silica gel, and a thermal conductivity detector. They exhibited the chromatograph at ACHEMA in Frankfurt, but nobody was interested in it.[10] N.C. Turner with the Burrell Corporation introduced in 1943 a massive instrument that used a charcoal column and mercury vapors. Stig Claesson of Uppsala University published in 1946 his work on a charcoal column that also used mercury.[10] Gerhard Hesse, while a professor at the University of Marburg/Lahn decided to test the prevailing opinion among German chemists that molecules could not be separated in a moving gas stream. He set up a simple glass column filled with starch and successfully separated bromine and iodine using nitrogen as the carrier gas. He then built a system that flowed an inert gas through a glass condenser packed with silica gel and collected the eluted fractions.[10] Courtenay S.G Phillips of Oxford University investigated separation in a charcoal column using a thermal conductivity detector. He consulted with Claesson and decided to use displacement as his separating principle. After learning about the results of James and Martin, he switched to partition chromatography.[10]

Column technology

[edit]

Early gas chromatography used packed columns, made of block 1–5 m long, 1–5 mm diameter, and filled with particles. The resolution of packed columns was improved by the invention of capillary column, in which the stationary phase is coated on the inner wall of the capillary.[6]

Physical components

[edit]

Autosamplers

[edit]

The autosampler provides the means to introduce a sample automatically into the inlets. Manual insertion of the sample is possible but is no longer common. Automatic insertion provides better reproducibility and time-optimization.

An autosampler for liquid or gaseous samples based on a microsyringe
An autosampler for liquid or gaseous samples based on a microsyringe

Different kinds of autosamplers exist. Autosamplers can be classified in relation to sample capacity (auto-injectors vs. autosamplers, where auto-injectors can work a small number of samples), to robotic technologies (XYZ robot[11] vs. rotating robot – the most common), or to analysis:

Inlets

[edit]
Split/splitless inlet

The column inlet (or injector) provides the means to introduce a sample into a continuous flow of carrier gas. The inlet is a piece of hardware attached to the column head.

Common inlet types are:

  • S/SL (split/splitless) injector – a sample is introduced into a heated small chamber via a syringe through a septum – the heat facilitates volatilization of the sample and sample matrix. The carrier gas then either sweeps the entirety (splitless mode) or a portion (split mode) of the sample into the column. In split mode, a part of the sample/carrier gas mixture in the injection chamber is exhausted through the split vent. Split injection is preferred when working with samples with high analyte concentrations (>0.1%) whereas splitless injection is best suited for trace analysis with low amounts of analytes (<0.01%). In splitless mode the split valve opens after a pre-set amount of time to purge heavier elements that would otherwise contaminate the system. This pre-set (splitless) time should be optimized, the shorter time (e.g., 0.2 min) ensures less tailing but loss in response, the longer time (2 min) increases tailing but also signal.[12]
  • On-column inlet – the sample is here introduced directly into the column in its entirety without heat, or at a temperature below the boiling point of the solvent. The low temperature condenses the sample into a narrow zone. The column and inlet can then be heated, releasing the sample into the gas phase. This ensures the lowest possible temperature for chromatography and keeps samples from decomposing above their boiling point.
  • PTV injector – Temperature-programmed sample introduction was first described by Vogt in 1979.[citation needed] Originally Vogt developed the technique as a method for the introduction of large sample volumes (up to 250 μL) in capillary GC. Vogt introduced the sample into the liner at a controlled injection rate. The temperature of the liner was chosen slightly below the boiling point of the solvent. The low-boiling solvent was continuously evaporated and vented through the split line. Based on this technique, Poy developed the programmed temperature vaporising injector; PTV. By introducing the sample at a low initial liner temperature many of the disadvantages of the classic hot injection techniques could be circumvented.[citation needed]
  • Gas source inlet or gas switching valve – gaseous samples in collection bottles are connected to what is most commonly a six-port switching valve. The carrier gas flow is not interrupted while a sample can be expanded into a previously evacuated sample loop. Upon switching, the contents of the sample loop are inserted into the carrier gas stream.
  • P/T (purge-and-trap) system – An inert gas is bubbled through an aqueous sample causing insoluble volatile chemicals to be purged from the matrix. The volatiles are 'trapped' on an absorbent column (known as a trap or concentrator) at ambient temperature. The trap is then heated and the volatiles are directed into the carrier gas stream. Samples requiring preconcentration or purification can be introduced via such a system, usually hooked up to the S/SL port.

The choice of carrier gas (mobile phase) is important. Hydrogen has a range of flow rates that are comparable to helium in efficiency. However, helium may be more efficient and provide the best separation if flow rates are optimized. Helium is non-flammable and works with a greater number of detectors and older instruments. Therefore, helium is the most common carrier gas used. However, the price of helium has gone up considerably over recent years, causing an increasing number of chromatographers to switch to hydrogen gas. Historical use, rather than rational consideration, may contribute to the continued preferential use of helium.

Detectors

[edit]

Commonly used detectors are the flame ionization detector (FID) and the thermal conductivity detector (TCD). While TCDs are beneficial in that they are non-destructive, its low detection limit for most analytes inhibits widespread use.[1] FIDs are sensitive primarily to hydrocarbons, and are more sensitive to them than TCD.[4] FIDs cannot detect water or carbon dioxide which make them ideal for environmental organic analyte analysis.[1] FID is two to three times more sensitive to analyte detection than TCD.[1]

The TCD relies on the thermal conductivity of matter passing around a thin wire of tungsten-rhenium with a current traveling through it.[4] In this set up helium or nitrogen serve as the carrier gas because of their relatively high thermal conductivity which keep the filament cool and maintain uniform resistivity and electrical efficiency of the filament.[4][13] When analyte molecules elute from the column, mixed with carrier gas, the thermal conductivity decreases while there is an increase in filament temperature and resistivity resulting in fluctuations in voltage ultimately causing a detector response.[4][13] Detector sensitivity is proportional to filament current while it is inversely proportional to the immediate environmental temperature of that detector as well as flow rate of the carrier gas.[4]

In a flame ionization detector (FID), electrodes are placed adjacent to a flame fueled by hydrogen / air near the exit of the column, and when carbon containing compounds exit the column they are pyrolyzed by the flame.[4][13] This detector works only for organic / hydrocarbon containing compounds due to the ability of the carbons to form cations and electrons upon pyrolysis which generates a current between the electrodes.[4][13] The increase in current is translated and appears as a peak in a chromatogram. FIDs have low detection limits (a few picograms per second) but they are unable to generate ions from carbonyl containing carbons.[4] FID compatible carrier gasses include helium, hydrogen, nitrogen, and argon.[4][13]

In FID, sometimes the stream is modified before entering the detector. A methanizer converts carbon monoxide and carbon dioxide into methane so that it can be detected. A different technology is the polyarc, by Activated Research Inc, that converts all compounds to methane.

Alkali flame detector (AFD) or alkali flame ionization detector (AFID) has high sensitivity to nitrogen and phosphorus, similar to NPD. However, the alkaline metal ions are supplied with the hydrogen gas, rather than a bead above the flame. For this reason AFD does not suffer the "fatigue" of the NPD, but provides a constant sensitivity over long period of time. In addition, when alkali ions are not added to the flame, AFD operates like a standard FID. A catalytic combustion detector (CCD) measures combustible hydrocarbons and hydrogen. Discharge ionization detector (DID) uses a high-voltage electric discharge to produce ions.

Flame photometric detector (FPD) uses a photomultiplier tube to detect spectral lines of the compounds as they are burned in a flame. Compounds eluting off the column are carried into a hydrogen fueled flame which excites specific elements in the molecules, and the excited elements (P,S, Halogens, Some Metals) emit light of specific characteristic wavelengths.[13] The emitted light is filtered and detected by a photomultiplier tube.[4][13] In particular, phosphorus emission is around 510–536 nm and sulfur emission is at 394 nm.[4][13] With an atomic emission detector (AED), a sample eluting from a column enters a chamber which is energized by microwaves that induce a plasma.[13] The plasma causes the analyte sample to decompose and certain elements generate an atomic emission spectra.[13] The atomic emission spectra is diffracted by a diffraction grating and detected by a series of photomultiplier tubes or photo diodes.[13]

Electron capture detector (ECD) uses a radioactive beta particle (electron) source to measure the degree of electron capture. ECD are used for the detection of molecules containing electronegative / withdrawing elements and functional groups like halogens, carbonyl, nitriles, nitro groups, and organometalics.[4][13] In this type of detector either nitrogen or 5% methane in argon is used as the mobile phase carrier gas.[4][13] The carrier gas passes between two electrodes placed at the end of the column, and adjacent to the cathode (negative electrode) resides a radioactive foil such as 63Ni.[4][13] The radioactive foil emits a beta particle (electron) which collides with and ionizes the carrier gas to generate more ions resulting in a current.[4][13] When analyte molecules with electronegative / withdrawing elements or functional groups electrons are captured which results in a decrease in current generating a detector response.[4][13]

Nitrogen–phosphorus detector (NPD), a form of thermionic detector where nitrogen and phosphorus alter the work function on a specially coated bead and a resulting current is measured.

Dry electrolytic conductivity detector (DELCD) uses an air phase and high temperature (v. Coulsen) to measure chlorinated compounds.

Mass spectrometer (MS), also called GC-MS; highly effective and sensitive, even in a small quantity of sample. This detector can be used to identify the analytes in chromatograms by their mass spectrum.[14] Some GC-MS are connected to an NMR spectrometer which acts as a backup detector. This combination is known as GC-MS-NMR.[citation needed] Some GC-MS-NMR are connected to an infrared spectrophotometer which acts as a backup detector. This combination is known as GC-MS-NMR-IR. It must, however, be stressed this is very rare as most analyses needed can be concluded via purely GC-MS.[citation needed]

Vacuum ultraviolet (VUV) represents the most recent development in gas chromatography detectors. Most chemical species absorb and have unique gas phase absorption cross sections in the approximately 120–240 nm VUV wavelength range monitored. Where absorption cross sections are known for analytes, the VUV detector is capable of absolute determination (without calibration) of the number of molecules present in the flow cell in the absence of chemical interferences.[15]

Olfactometric detector, also called GC-O, uses a human assessor to analyse the odour activity of compounds. With an odour port or a sniffing port, the quality of the odour, the intensity of the odour and the duration of the odour activity of a compound can be assessed.

Other detectors include the Hall electrolytic conductivity detector (ElCD), helium ionization detector (HID), infrared detector (IRD), photo-ionization detector (PID), pulsed discharge ionization detector (PDD), and thermionic ionization detector (TID).[16]

Methods

[edit]
This image above shows the interior of a GeoStrata Technologies Eclipse Gas Chromatograph that runs continuously in three-minute cycles. Two valves are used to switch the test gas into the sample loop. After filling the sample loop with test gas, the valves are switched again applying carrier gas pressure to the sample loop and forcing the sample through the column for separation.

The method is the collection of conditions in which the GC operates for a given analysis. Method development is the process of determining what conditions are adequate and/or ideal for the analysis required.

Conditions which can be varied to accommodate a required analysis include inlet temperature, detector temperature, column temperature and temperature program, carrier gas and carrier gas flow rates, the column's stationary phase, diameter and length, inlet type and flow rates, sample size and injection technique. Depending on the detector(s) (see below) installed on the GC, there may be a number of detector conditions that can also be varied. Some GCs also include valves which can change the route of sample and carrier flow. The timing of the opening and closing of these valves can be important to method development.

Carrier gas selection and flow rates

[edit]

Typical carrier gases include helium, nitrogen, argon, and hydrogen.[4][1] Which gas to use is usually determined by the detector being used, for example, a DID requires helium as the carrier gas.[1] When analyzing gas samples the carrier is also selected based on the sample's matrix, for example, when analyzing a mixture in argon, an argon carrier is preferred because the argon in the sample does not show up on the chromatogram. Safety and availability can also influence carrier selection.

The purity of the carrier gas is also frequently determined by the detector, though the level of sensitivity needed can also play a significant role. Typically, purities of 99.995% or higher are used. The most common purity grades required by modern instruments for the majority of sensitivities are 5.0 grades, or 99.999% pure meaning that there is a total of 10 ppm of impurities in the carrier gas that could affect the results. The highest purity grades in common use are 6.0 grades, but the need for detection at very low levels in some forensic and environmental applications has driven the need for carrier gases at 7.0 grade purity and these are now commercially available. Trade names for typical purities include "Zero Grade", "Ultra-High Purity (UHP) Grade", "4.5 Grade" and "5.0 Grade".

The carrier gas linear velocity affects the analysis in the same way that temperature does (see above). The higher the linear velocity the faster the analysis, but the lower the separation between analytes. Selecting the linear velocity is therefore the same compromise between the level of separation and length of analysis as selecting the column temperature. The linear velocity will be implemented by means of the carrier gas flow rate, with regards to the inner diameter of the column.

With GCs made before the 1990s, carrier flow rate was controlled indirectly by controlling the carrier inlet pressure, or "column head pressure". The actual flow rate was measured at the outlet of the column or the detector with an electronic flow meter, or a bubble flow meter, and could be an involved, time consuming, and frustrating process. It was not possible to vary the pressure setting during the run, and thus the flow was essentially constant during the analysis. The relation between flow rate and inlet pressure is calculated with Poiseuille's equation for compressible fluids.

Many modern GCs, however, electronically measure the flow rate, and electronically control the carrier gas pressure to set the flow rate. Consequently, carrier pressures and flow rates can be adjusted during the run, creating pressure/flow programs similar to temperature programs.

Stationary compound selection

[edit]

The polarity of the solute is crucial for the choice of stationary compound, which in an optimal case would have a similar polarity as the solute. Common stationary phases in open tubular columns are cyanopropylphenyl dimethyl polysiloxane, carbowax polyethyleneglycol, biscyanopropyl cyanopropylphenyl polysiloxane and diphenyl dimethyl polysiloxane. For packed columns more options are available.[4]

Inlet types and flow rates

[edit]

The choice of inlet type and injection technique depends on if the sample is in liquid, gas, adsorbed, or solid form, and on whether a solvent matrix is present that has to be vaporized. Dissolved samples can be introduced directly onto the column via a COC injector, if the conditions are well known; if a solvent matrix has to be vaporized and partially removed, a S/SL injector is used (most common injection technique); gaseous samples (e.g., air cylinders) are usually injected using a gas switching valve system; adsorbed samples (e.g., on adsorbent tubes) are introduced using either an external (on-line or off-line) desorption apparatus such as a purge-and-trap system, or are desorbed in the injector (SPME applications).

Sample size and injection technique

[edit]

Sample injection

[edit]
The rule of ten in gas chromatography

The real chromatographic analysis starts with the introduction of the sample onto the column. The development of capillary gas chromatography resulted in many practical problems with the injection technique. The technique of on-column injection, often used with packed columns, is usually not possible with capillary columns. In the injection system in the capillary gas chromatograph the amount injected should not overload the column and the width of the injected plug should be small compared to the spreading due to the chromatographic process. Failure to comply with this latter requirement will reduce the separation capability of the column. As a general rule, the volume injected, Vinj, and the volume of the detector cell, Vdet, should be about 1/10 of the volume occupied by the portion of sample containing the molecules of interest (analytes) when they exit the column.

Some general requirements which a good injection technique should fulfill are that it should be possible to obtain the column's optimum separation efficiency, it should allow accurate and reproducible injections of small amounts of representative samples, it should induce no change in sample composition, it should not exhibit discrimination based on differences in boiling point, polarity, concentration or thermal/catalytic stability, and it should be applicable for trace analysis as well as for undiluted samples.

However, there are a number of problems inherent in the use of syringes for injection. Even the best syringes claim an accuracy of only 3%, and in unskilled hands, errors are much larger. The needle may cut small pieces of rubber from the septum as it injects sample through it. These can block the needle and prevent the syringe filling the next time it is used. It may not be obvious that this has happened. A fraction of the sample may get trapped in the rubber, to be released during subsequent injections. This can give rise to ghost peaks in the chromatogram. There may be selective loss of the more volatile components of the sample by evaporation from the tip of the needle.[17]

Column selection

[edit]

The choice of column depends on the sample and the active measured. The main chemical attribute regarded when choosing a column is the polarity of the mixture, but functional groups can play a large part in column selection. The polarity of the sample must closely match the polarity of the column stationary phase to increase resolution and separation while reducing run time. The separation and run time also depends on the film thickness (of the stationary phase), the column diameter and the column length.

Column temperature and temperature program

[edit]
A gas chromatography oven, open to show a capillary column

The column(s) in a GC are contained in an oven, the temperature of which is precisely controlled electronically. (When discussing the "temperature of the column," an analyst is technically referring to the temperature of the column oven. The distinction, however, is not important and will not subsequently be made in this article.)

The rate at which a sample passes through the column is directly proportional to the temperature of the column. The higher the column temperature, the faster the sample moves through the column. However, the faster a sample moves through the column, the less it interacts with the stationary phase, and the less the analytes are separated.

In general, the column temperature is selected to compromise between the length of the analysis and the level of separation.

A method which holds the column at the same temperature for the entire analysis is called "isothermal". Most methods, however, increase the column temperature during the analysis, the initial temperature, rate of temperature increase (the temperature "ramp"), and final temperature are called the temperature program.

A temperature program allows analytes that elute early in the analysis to separate adequately, while shortening the time it takes for late-eluting analytes to pass through the column.

Data reduction and analysis

[edit]

Qualitative analysis

[edit]

Generally, chromatographic data is presented as a graph of detector response (y-axis) against retention time (x-axis), which is called a chromatogram. This provides a spectrum of peaks for a sample representing the analytes present in a sample eluting from the column at different times. Retention time can be used to identify analytes if the method conditions are constant. Also, the pattern of peaks will be constant for a sample under constant conditions and can identify complex mixtures of analytes. However, in most modern applications, the GC is connected to a mass spectrometer or similar detector that is capable of identifying the analytes represented by the peaks.

Quantitative analysis

[edit]

The area under a peak is proportional to the amount of analyte present in the chromatogram. By calculating the area of the peak using the mathematical function of integration, the concentration of an analyte in the original sample can be determined. Concentration can be calculated using a calibration curve created by finding the response for a series of concentrations of analyte, or by determining the relative response factor of an analyte. The relative response factor is the expected ratio of an analyte to an internal standard (or external standard) and is calculated by finding the response of a known amount of analyte and a constant amount of internal standard (a chemical added to the sample at a constant concentration, with a distinct retention time to the analyte).

In most modern GC-MS systems, computer software is used to draw and integrate peaks, and match MS spectra to library spectra.

Applications

[edit]

In general, substances that vaporize below 300 °C (and therefore are stable up to that temperature) can be measured quantitatively. The samples are also required to be salt-free; they should not contain ions. Very minute amounts of a substance can be measured, but it is often required that the sample must be measured in comparison to a sample containing the pure, suspected substance known as a reference standard.

Various temperature programs can be used to make the readings more meaningful; for example to differentiate between substances that behave similarly during the GC process.

Professionals working with GC analyze the content of a chemical product, for example in assuring the quality of products in the chemical industry; or measuring chemicals in soil, air or water, such as soil gases.[18] GC is very accurate if used properly and can measure picomoles of a substance in a 1 ml liquid sample, or parts-per-billion concentrations in gaseous samples.

In practical courses at colleges, students sometimes get acquainted to the GC by studying the contents of lavender oil or measuring the ethylene that is secreted by Nicotiana benthamiana plants after artificially injuring their leaves. These GC analyse hydrocarbons (C2-C40+). In a typical experiment, a packed column is used to separate the light gases, which are then detected with a TCD. The hydrocarbons are separated using a capillary column and detected with a FID. A complication with light gas analyses that include H2 is that He, which is the most common and most sensitive inert carrier (sensitivity is proportional to molecular mass) has an almost identical thermal conductivity to hydrogen (it is the difference in thermal conductivity between two separate filaments in a Wheatstone Bridge type arrangement that shows when a component has been eluted). For this reason, dual TCD instruments used with a separate channel for hydrogen that uses nitrogen as a carrier are common. Argon is often used when analysing gas phase chemistry reactions such as F-T synthesis so that a single carrier gas can be used rather than two separate ones. The sensitivity is reduced, but this is a trade off for simplicity in the gas supply.

Gas chromatography is used extensively in forensic science. Disciplines as diverse as solid drug dose (pre-consumption form) identification and quantification, arson investigation, paint chip analysis, and toxicology cases, employ GC to identify and quantify various biological specimens and crime-scene evidence.

See also

[edit]

References

[edit]
  1. ^ a b c d e f g h i Harvey, David (2000). Modern analytical chemistry. Boston: McGraw-Hill. ISBN 0-07-237547-7. OCLC 41070677.
  2. ^ a b Pavia, L.; Gary M. Lampman; George S. Kritz; Randall G. Engel (2006). Introduction to Organic Laboratory Techniques (4th ed.). Thomson Brooks/Cole. pp. 797–817. ISBN 978-0-495-28069-9.
  3. ^ "Gas Chromatography". Linde AG. Archived from the original on 3 March 2012. Retrieved 11 March 2012.
  4. ^ a b c d e f g h i j k l m n o p q r s Harris, Daniel C.; Charles A. Lucy (2016). Quantitative chemical analysis (Ninth ed.). New York: W.H. Freeman & Company. ISBN 978-1-4641-3538-5. OCLC 915084423.
  5. ^ "Berichte der Deutschen Botanischen Gesellschaft v.24 1906". HathiTrust. 1883. Retrieved 2019-04-19.
  6. ^ a b Bartle, Keith D.; Myers, Peter (10 September 2002). "History of gas chromatography". TrAC Trends in Analytical Chemistry. 21 (9): 547–557. doi:10.1016/S0165-9936(02)00806-3.
  7. ^ James, A. T.; Martin, A. J. P. (1 March 1952). "Gas-liquid partition chromatography: the separation and micro-estimation of volatile fatty acids from formic acid to dodecanoic acid". Biochemical Journal. 50 (5): 679–690. doi:10.1042/bj0500679. PMC 1197726. PMID 14934673.
  8. ^ R. A. Dewar; McWilliam, I. G. (March 1958). "Flame Ionization Detector for Gas Chromatography". Nature. 181 (4611): 760. Bibcode:1958Natur.181..760M. doi:10.1038/181760a0. ISSN 1476-4687. S2CID 4175977.
  9. ^ Martin, A. J. P.; Synge, R. L. M. (1 December 1941). "A new form of chromatogram employing two liquid phases: A theory of chromatography. 2. Application to the micro-determination of the higher monoamino-acids in proteins". Biochemical Journal. 35 (12): 1358–1368. doi:10.1042/bj0351358. PMC 1265645. PMID 16747422.
  10. ^ a b c d Ettre, Leslie S. (2008). Chapters in the evolution of chromatography. London: Imperial College Press. ISBN 978-1860949432.
  11. ^ Carvalho, Matheus (2018). "Osmar, the open-source microsyringe autosampler". HardwareX. 3: 10–38. doi:10.1016/j.ohx.2018.01.001.
  12. ^ Chasteen, Thomas G. "Split/Splitless and On-Column Gas Chromatographic Injectors". Retrieved October 6, 2019.
  13. ^ a b c d e f g h i j k l m n o p Higson, S. (2004). Analytical Chemistry. OXFORD University Press ISBN 978-0-19-850289-0
  14. ^ Skoog, Douglas A.; West, Donald M.; James Holler, F.; Crouch, Stanley R. (2013-01-01). Skoog, Douglas A.; West, Donald M.; Holler, F. James; Crouch, Stanley R. (eds.). Fundamentals of analytical chemistry (Ninth ed.). Belmont, CA: Brooks/Cole. ISBN 9780495558286. OCLC 824171785.
  15. ^ Schug, Kevin A.; Sawicki, Ian; Carlton, Doug D.; Fan, Hui; McNair, Harold M.; Nimmo, John P.; Kroll, Peter; Smuts, Jonathan; Walsh, Phillip; Harrison, Dale (1834). "Vacuum Ultraviolet Detector for Gas Chromatography". Analytical Chemistry. 86 (16): 8329–35. doi:10.1021/ac5018343. PMID 25079505.
  16. ^ Poole, Colin F. (2015-11-20). "Ionization-based detectors for gas chromatography". Journal of Chromatography A. 1421: 137–153. doi:10.1016/j.chroma.2015.02.061. PMID 25757823.
  17. ^ Grob, Robert L.; Barry, Eugene F. (2004). Modern Practice of Gas Chromatography (4th Ed.). John Wiley & Sons. ISBN 978-0-471-22983-4.
  18. ^ Kim, D; Vargas, R; Bond-Lamberty, B; Turetsky, M (2012). "Effects of soil rewetting and thawing on soil gas fluxes: a review of current literature and suggestions for future research". Biogeosciences. 9 (7): 3459–3483. Bibcode:2012BGeo....9.2459K. doi:10.5194/bg-9-2459-2012.
[edit]

Media related to Gas chromatography at Wikimedia Commons